04-MAY-2011 Bug fix Problem with starting View Data from the Formation Shell addressed.
MacTest32 board configuration: Going to 1 chn per board did not update the board type correctly.
11-MAY-2011 Bug fix Resuming a test channel caused end on cycle to terminate
18-MAY-2011 Bug fix In case of updating SW and extra program folder was created in the start menu.
20-MAY-2011 Bug fix Digital input data were not reported in the data file
20-MAY-2011 Bug fix Error trapping of errors in the user created DLL's
23-MAY-2011 Bug fix The rate of communication slowed dramatically when pulling down menus, dialogs present or Alt
key pressed. Fixed.
25-MAY-2011 New feature SW generated ends are delayed 10 ticks into steps.
Special_GUI=Digital_IO_1_3 added.
24-MAY-2011 Bug fix "Day 1" problem with popped up dialogs, pulled down menus etc. slowing down communication fixed.
01-JUN-2011 New feature MacNotify added. This feature will send emails on selected events.
SystemType 5 added for Maccor designed 8500 brain board.
02-JUN-2011 Bug fix Problem when Build Test auto-upgrades certain EXTERNAL DLL data in the comment field to the new
procedure format.
06-JUN-2011 Bug fix Logged error traps added to all timers. Some additional details will go in new WinEvent.txt file.
08-JUN-2011 Bug fix Automatic SMB scan list refresh added.
09-JUN-2011 Bug fix Version control of downloaded binaries at program start. If they are different a board reset will
be requested.
Special binary versions are displayed clearly on the main screen.
15-JUN-2011 Bug fix Incorrect setting of variables at end of WF steps fixed.
21-JUN-2011 New feature SystemType 5 changed to be single range.
22-JUN-2011 Bug fix If a double quote " was the last charater in the proc. comment, the procedure would be corrupt.
23-JUN-2011 New feature ProcessSafeties migrated to Delphi
12-JUL-2011 New feature Step time, Capacity and Energy optionally carried into and out of WF steps.
13-JUL-2011 Bug fix Error in capacity and energy values in functions during waveforms fixed
Error 164: Timeout during waveform could not be resumed - fixed.
01-AUG-2011 Bug fix Problem with some custom reports showing "Chinese" characters fixed.
02-AUG-2011 Bug fix Printing procedures as barcodes didn't work - fixed
Barcode operation didn't work from the main screen - fixed.
08-AUG-2011 New feature Support for CAN in functions added.
09-AUG-2011 Bug fix Procedures with EV steps in subroutines only were not detected in the start dialog - fixed.
05-SEP-2011 Bug fix Internal modification of the XML procedure format
The Jump To feature did not work correctly with new XML format procedure.
16-SEP-2011 New feature Support for environmental chambers added to 500 ms (5300) systems
Defrosting of Maccor environmental chambers changed to run on a timer
19-SEP-2011 New feature FW Update: Added: 12/14 bit mode detect for new SR 32 cell board. 50 ms mode
FW Update: Added: 12/14 bit mode detect for new SR 32 cell board. 500 ms mode
24-SEP-2011 Bug fix Improved power fail handling in 500 ms mode.
30-SEP-2011 Bug fix Autorange in waveforms on single, dual and triple range channels didn't work. Fixed.
19-OCT-2011 New feature CAN additions: Send CAN messages, Limit waveforms by CAN
Additional checks and logging of failing functions added
20-OCT-2011 Bug fix Crashed caused by sub routines fixed
16 end conditions could in some cases cause the reports to become ends.
24-OCT-2011 Bug fix The use of non-English characters in test procedures could cause problems. Fixed now.
05-NOV-2011 New feature UTF-8 tag removed from procedure files
Added checks for the validity of SMB scan lists
Removed several truncations of data from larger SMB scan lists
Maccor EV Chamber handling rewritten.
14-NOV-2011 Bug fix The Jump feature copied the step number to the operator. Fixed.
15-NOV-2011 Bug fix Display of Windows version in the About box corrected.
Error handlers with logging added to all MacTest32.dll calls
14-DEC-2011 Bug fix The formation shell did not start all channels. Fixed.
Additional crash logging added to MacTest32. Unhandled crashed will create or append to a
C:\Maccor\ExcMagic.Log file.
Multiple bug fixes in the Maccor Environmental Chamber handling.
Running old format test procedure on version 3 SW caused the Jump feature to lock up.
16-DEC-2011 Bug fix Pop up of mass entry box when running regimes fixed.
20-DEC-2011 Bug fix Memory leak when using function end conditions or IF in function steps fixed
16-JAN-2012 New feature SMB3 updated to support longer messages.
Maccor EV chamber updated for speed, stability and safety
Check added for long tests and +4 ends and reports in formation mode.
Send emails from test procedures.
17-JAN-2012 Bug fix Combined Triple Range issue fixed. Combined channel on formation formation issue fixed
18-JAN-2012 Release version 3.1
19-JAN-2012 Bug fix Multiple SMB3 boards and few connected batteries would cause the system to almost lock up.
01-FEB-2012 New feature Number of variables increased to 15
10-FEB-2012 New feature General audit feature added
"False available" checked for in audit. Will reset the board to recover
Available channels will occasionally send a ChStop message
File archiving updated. Better clean up when channels are archived or reset.
14-FEB-2012 New feature Environmental chamber type 24, ITC, added. Environmental chamber type 25, Thermo Scentific, added
Environmental chamber type 26, Binder MB1 (with program option), added.
Environmental chamber type 27, ESPEC SCP-220 with humidity control, added
24-FEB-2012 Bug fix Power fail on System Type 1 caused combined channels to uncombine - fixed
25-FEB-2012 New feature Separate settings for cool and heat for fan problem detection in Maccor env. chambers(22) added.
13-MAR-2012 New feature Check of file size and number of archived files added.
23-MAR-2012 New feature Variables added to the data files. Variables reloaded when a stopped test is restarted.
04-APR-2012 New feature ITC chamber (24) is now only setting the temperature when the channel is active
02-MAY-2012 New feature Master controller board communication code rewritten. End conditions in 5300 mode are now
processed by the SW.
"ProblemSuspendsAll" added. With this set, any channel that goes into problem will suspend all
other channels.
03-MAY-2012 Bug fix Brick power safety settings kept getting reset on large (+300 W) channels. Fixed.
Missing controller boards no longer slow down the ping rate.
06-MAY-2012 Bug fix Pulsing on combined channel prohibited.
14-MAY-2012 Bug fix Out of range function values are now trapped as problems.
25-MAY-2012 Bug fix Setting variables in the step before a sub routine didn't work. Fixed.
30-MAY-2012 New feature Size of SysEvents.log increased to 100.000 entries.
Release and Build numbers changed to Build DateTime.
07-JUN-2012 Bug fix Fixed memory leak in funtions.
18-JUN-2012 New feature Added support for EV chamber type 28: Thermotron 8800.
05-JUL-2012 New feature Added digital output to EV chambers. So far type 27 is supported.
Added digital input to EV chambers. Reading accessible in functions. So far type 27 is supported.
Added access to EV temperature and humidity in functions.
Added Proceed feature to EV chamber steps. This will cause the test to proceed as soon as the
temperature is set.
Added access to variables from other test channels.
Added 32 global flags to functions
11-JUL-2012 New feature Handle global safety end conditions from Build Test
17-JUL-2012 Bug fix Recover voltage readings when a channel is uncombined.
29-NOV-2012 New feature Added support for VIN board based thermistors with full Steihart-Hart linearization.
30-NOV-2012 New feature SMB configuration import/export added.
30-JAN-2013 Bug fix Erroneous times in Advanced Start -> Append to archived fixed.
01-MAR-2013 Bug fix Problems with restarting archived tests fixed. Truncation wasn't right and current and voltage
could be blanked out.
06-MAR-2013 New feature Added support for EV Chamber 30: TEMI 880 PCL1.
Added support for EV Chamber 31: Generic MODBUS/TCP/IP
14-MAR-2013 Bug fix Problems with false positive audits addressed.
23-MAR-2013 New feature Multiple 8500 updates, Ranging glitches, Max rate up to 1 kHz data rate added,
Down to 90 ms chg/dis switch time option added. NO FRA/EDA support in this version.
06-MAY-2013 Bug fix Problem with end conditions in formation mode fixed.
24-MAY-2013 Bug fix Several issues with end conditions and reports fixed in 5300 mode.
28-MAY-2013 New feature Functions added to pulsing.
03-JUN-2013 New feature Support added for thermistor boards with 10.7 K parallel resistor.
13-JUN-2013 Bug fix Problem with thermistor temperatures being displayed and used incorrectly in function ends in
some cases corrected.
18-JUN-2013 New feature Functions added to pulse mode.
05-JUL-2013 New feature Hyperswitch mode added
19-JUL-2013 New feature FRA handling rewritten. Support for Solartron 1260/1287 added. Support for ParSTAT 4000 added.
13-AUG-2013 Bug fix Audit modifications: False available only checked in 500 ms mode, Wrong test time check removed,
Audit events pop up
14-AUG-2013 New feature 500 ms mode can now run with 0 V
Environmental chambers added to initialization screen 5
15-AUG-2013 New feature Bitwise AND, OR and NOT functions added
500 ms mode can now run with Max V
26-AUG-2013 New feature Support for KEBA UMC1100 and UMC1200 environmental chamber controllers added (31/1/1)
20-SEP-2013 New feature Support for digital input via aux boards added.
11-OCT-2013 New feature Manual FRA mux override added.
KEBA UMC1100 and UMC1200 environmental chambers can optionally be started and stopped from the SW
29-OCT-2013 Bug fix Problem with sporadically missed function ends fixed.
01-NOV-2013 Bug fix Better recovery of tests not recording data.
05-NOV-2013 New feature Support for the EDA reestablished.
12-NOV-2013 New feature The calibration in MacTest32 can now use an Agilent 34410A multimeter.
14-NOV-2013 New feature Support for Solartron 1255B/1287 added.
09-JAN-2014 New feature EV Chamber 34 added: HANYOUG TD500 PC-Link
EV Chamber 35
14-JAN-2014 Bug fix Minus (-) operator for auxiliaries fixed.
01-FEB-2014 New feature More than 48 aux per channel. Same aux can be assigned to more than one test channel.
New aux assigment screen.
12-FEB-2014 Bug fix Delay for SW handled current end conditions in 500 ms mode, so it must be met 3 times in a row
before it is fired.
14-FEB-2014 Bug fix Pulse rules updated. Max 3 out of 4 channels in 1 ms pulse mode
Barcode operation trimmed
20-FEB-2014 Bug fix Current regulation added for 8500
23-FEB-2014 Bug fix Fix of "Inhibit step time reset" in 5300 mode
25-FEB-2014 Bug fix Setting variables in EV steps could cause a crash. Fixed.
03-MAR-2014 Bug fix Misc updates of ParSTAT 4000 and Solartron 1287/1260
27-MAR-2014 New feature Support for Kepco booster added.
EV Chamber 36 added: HANYOUG TH500 PC-Link
03-APR-2014 Bug fix Reverse byte order words added to SMB (Word Flip)
Global flags only got immediately updated in the file of the channel that changed it - fixed.
13-APR-2014 Bug fix Dual Range \ MultiRange configuration conflict fixed
24-APR-2014 Bug fix Global flags not getting updated correctly when entering and leaving subroutines
30-APR-2014 New feature EV Chamber 29 added: Cincinnati Sub-Zero EZT-570i.
01-MAY-2014 New feature EV Chamber 37 added: Cincinnati Sub-Zero EZT-430i.
05-MAY-2014 Bug fix Lock added to prevent erroneous jumping into or out of pause steps.
15-MAY-2014 New feature BIA Climatic environmental chamber added as type 38.
20-MAY-2014 Bug fix Pulse durations not being tick multiples were not trapped as errors in all cases.
21-MAY-2014 Bug fix Could not assign AUX channels from second AUX board fixed.
22-MAY-2014 Bug fix Digital output only worked on the first AUX board fixed.
23-MAY-2014 New feature TxType 4, SubType 7 and 8 added to MacNet.
28-MAY-2014 Bug fix Problems with assigning the last aux on a panel fixed.
06-JUN-2014 Bug fix Lock added to prevent a pulse in step 1 added.
19-JUN-2014 Bug fix Extended CAN messages fixed.
27-JUN-2014 Bug fix Saving the channel/board assignment caused an index out of boundary error. Fixed.
03-JUL-2014 New feature Support for Modulab FRA added.
29-JUL-2014 Bug fix Multi-Level-Pulse rules updated.
06-AUG-2014 Bug fix Problem with EDA mux not switching fixed.
Added logging of SMR, Jump, and restart.
06-AUG-2014 Bug fix Multi Level Pulse will not recover from: Problem/Clear Channel/Advance Start to MLP. (FW)
15-AUG-2014 Bug fix Problem with procedure in the detailed screen not updating after new header fixed.
18-AUG-2014 Bug fix Verification of downloaded calibration added to 500 ms mode (5300)
25-AUG-2014 Bug fix Redundant copy of range scale factors removed.
27-AUG-2014 Bug fix Lock added to prevent advanced starting into certain step types added.
01-SEP-2014 New feature Environmental_Chamber_Optional_String added to EV chamber type 17
05-SEP-2014 Bug fix Added consecutive read error count to HDQ8 packs to filter out random errors in data. That change
is in the SMB3 Client.
Filter on/off controlled by user via PC. (FW)
11-SEP-2014 Bug fix Digital output from formation shell not working - fixed
12-SEP-2014 Bug fix Wrong barcode entry of test procedure in start dialog fixed.
23-SEP-2014 Bug fix Suspend-Resume of FRA steps (re)added
21-OCT-2014 New feature Barcode option 3 and XML tray files added to the Formation Shell
22-OCT-2014 Bug fix Several barcode "adjustments" in the tester program
29-OCT-2014 New feature Thermocouple board verification feature added
31-OCT-2014 Bug fix Too long waveform file names were not handled correctly.
04-NOV-2014 Bug fix Several errors in CAN handling fixed including big endian.
06-NOV-2014 New feature VersaSTATwrapper.dll now loaded dynamically.
12-NOV-2014 New feature Restart archived test has been replaced by Advanced Start in the regular start dialog
15-NOV-2014 Bug fix Problem with delayed SW handled digital output fixed. (SW+FW)
08-DEC-2014 Bug fix Barcode handling optimized in Formation Shell XML Tray files added.
16-DEC-2014 New feature Support for environmental chambers added to regimes
19-DEC-2014 Bug fix Auto pulse rules updated
20-DEC-2014 New feature Manual pulse mode selection added
27-DEC-2014 New feature Start from file added
Direct control mode added to MacNet
Exit added to regimes
19-JAN-2015 Bug fix Lab systems are forced to range 1 in rest steps
27-JAN-2015 Bug fix Start from file fine tuned.
17-FEB-2015 Bug fix Relay switch timing default values set.
26-FEB-2015 Bug fix MacNet function (7, 4) read SMB fixed.
13-MAR-2015 New feature Addition of MacNet examples for network and LabView
09-APR-2015 New feature Support for Solartron 1260/1470 added.
13-APR-2015 Bug fix Incorrect legal ranges in maintenance and simple cal for some configurations addressed (M273)
15-APR-2015 New feature Data suppression in Waveforms added.
17-APR-2015 Bug fix Logging of power fail time out added (M274)
23-APR-2015 Bug fix SW Dig out being reset on subroutine entry and exit fixed.
29-APR-2015 Bug fix Problem with starting procedures with sub routines locking up the SW fixed.
12-MAY-2015 New feature Channel numbering offset added
Initialization screen 1 migrated
28-MAY-2015 New feature Support for 4000H brick added.
03-JUN-2015 Bug fix Missing ability to edit aux descriptions added
24-JUN-2015 New feature Optional addition of relay switch time to test time added. (SW + FW)
25-JUN-2015 Bug fix Problem with limits in MacNet direct mode fixed
08-JUL-2015 Bug fix Auxiliary reports did not work correctly - fixed.
09-JUL-2015 New feature UNIQUE UMC1100 environmental chamber added as type 40.
22-JUL-2015 Bug fix Door interlocks and smoke alarms rewritten.
Occasional AUX FW download lock-up fixed.
09-SEP-2015 New feature Support for 8500-IGBT added
18-SEP-2015 New feature Support for voltage divider added.
22-SEP-2015 Bug fix MacNet: Minor fixes and a watchdog timer added to the direct mode.
05-OCT-2015 Bug fix Restart into a function step blocked.
09-OCT-2015 New feature CAN write added.
13-OCT-2015 New feature Maccor EV chamber mark II added as chamber type 50
28-OCT-2015 Bug fix AUXTR and AUXTT did not work in functions: fixed.
02-NOV-2015 Bug fix Constant current steps with no limits but with voltage ends, get this end set as a limit.
SystemType 0 and 3 only
04-NOV-2015 New feature TATO TT-5166 environmental chamber added as type 39.
Settings VARs to Min, Max and continuously added.
06-NOV-2015 New feature Existing archived filename is now reported in start dialog.
Test proc. defined end condition by setting the goto to the step itself in function ends added.
07-NOV-2015 Bug fix Better clean up of TestPro files on failed test start.
10-NOV-2015 Bug fix Starting a test with C-rates without entering the C-rate is now reported in start dialog.
18-NOV-2015 New feature Added Mass to file based remote control
28-NOV-2015 Bug fix Errors with External Functions (DLL's) and subroutines fixed.
16-DEC-2015 Bug fix Bug in inserting test steps introduced NOV.6,2015 fixed.
18-DEC-2015 Bug fix Capacity calculation from Parallel R failed 596.5 hours after PC start. Fixed.
04-JAN-2016 Bug fix AC-impedance no longer licensed for system type 1
08-JAN-2016 New feature MacNet (7, 13) added. Read SMB header from battery.
11-JAN-2016 New feature MacNet (7, 14) and (7, 15) added. Get and set SMB position setup.
12-JAN-2016 New feature MacNet (4, 9) added. Get test time of multiple channels.
18-JAN-2016 New feature Change BoardConfig form to display settings as read-only. (Moved to Setup & Calibration.)[RH]
19-JAN-2016 Bug fix Fix ISO_CONN edit box showing even when Panel1 is not visible.
27-JAN-2016 Bug fix Erroneous request for C-rate entry fixed.
10-FEB-2016 New feature Ethernet support added to EV Chamber type 36
12-FEB-2016 Bug fix Erroneous pop up when jumping a test with C rate fixed.
16-FEB-2016 Bug fix EVChamber type 50 version control updated.
21-FEB-2016 Bug fix Add declarations for procedures SendProg and BuildChannelMsg to BasicLAN.
22-FEB-2016 New feature Add CheckChan_NEW (migrating). Remove unused Combine Chans tab since combined chans
accessed using Check Channels form. [RH]
26-FEB-2016 New feature Password-protect editing of no. Active Channels, Board Speed and Pulse Mode on (migrated)
Board Config form and Check Channels protected edits. [RH]
03-MAR-2016 New feature WF support added to SystemType 6
04-MAR-2016 New feature Setup net adapter migrated
04-MAR-2016 Bug fix Make Set, Copy Previous Channel, and Copy to Board more uniform.[RH]
05-MAR-2016 New feature Changing channel will no longer "save" channel specs. Only safeties can be set, and only when
Set is clicked; HW-related chn specs must now be changed from Calibration program.[RH]
07-MAR-2016 New feature Delete archived files migrated
09-MAR-2016 Bug fix EV Chamber type 40 problem with lower temperatures fixed.
10-MAR-2016 New feature Remove Reset button and Suspend from new Check Channels form.[RH]
16-MAR-2016 Bug fix Restore special device combo box to Board Config form.[RH]
21-MAR-2016 Bug fix Add missing ELC min/max voltage fields to CheckChan form. Require combining channels
to have same Vspecs. [RH]
29-MAR-2016 Bug fix Change min. board speed for 1 high-speed chn from 1 to 2ms. Add check of board speed.[RH]
31-MAR-2016 Bug fix Added check for "." in regime date file names.
Check for invalid channel numbers added to the Aux Assign screen.
06-APR-2016 Bug fix Always open Check Chan form on chn 1, Charge, Range 4, Constant Current, set Value of 0.
Make sure channel is stopped before changing channels. [RH]
21-APR-2016 Bug fix Board Timing (relay cycle time) changed from a system setting to a board setting.[RH]
12-MAY-2016 New feature EV Chambers 1-17 and 19-21 migrated to Delphi
17-MAY-2016 Bug fix Fixed Windows version numbers above 8.0 in debug info and about box.
18-MAY-2016 Bug fix Suspend-resume could cause Set Variable at End to be executed. Fixed.
19-MAY-2016 Bug fix Fix system going No Cntrller when Active chans set to 1 using new Board Config form. [RH]
19-MAY-2016 Bug fix Faster check of voltage and current safeties.
20-MAY-2016 Bug fix Verify good values entered on Check Channel form. Allow Low Current Control Delay to be
password-accessible.[RH]
Problems with multiple limits in MacNet direct mode fixed[JM]
20-MAY-2016 New feature Board Timing (relay cycle time) is now displayed in true time (seconds).
23-MAY-2016 Bug fix Add ILimit values for system types 1 & 3.
24-MAY-2016 Bug fix Board ping timing improved.
27-MAY-2016 New feature Authentication added to MacNet (8, 1) and (8, 2).
Check if a procedure can be started on a channel added to MacNet (6, 11) and (6, 12)
02-JUN-2016 New feature Advanced start added to MacNet (6, 13)
06-JUN-2016 New feature Added (1,3) Procedure Exists, (1, 4) Test File Name Exists, (1, 5) Get File Listing to MacNet.
08-JUN-2016 Bug fix Pulse rules updated. Must be preceded by a charge or discharge step of same range and
mode as the first level of the pulse. Minimum time in a range increased to 600 us.
09-JUN-2016 New feature Current end condition in waveforms included in auto-range select.
10-JUN-2016 New feature Configurable 8500 programming port added.
15-JUN-2016 Bug fix Fatal problem when SMR a test with C-rate fixed.
20-JUN-2016 New feature Pause steps are now firing a Suspend event.
22-JUN-2016 New feature MacNet functions (1, 6) and (1, 7) added to send and get files from the tester.
23-JUN-2016 Bug fix Download of SMB Clients improved.
27-JUN-2016 Bug fix Add SMB3 server timing. SetSMB, SetSMBext and SetCAN caused additional data points fixed.
28-JUN-2016 New feature Add Keysight 34465A meter to SimpleCal.[RH]
30-JUN-2016 New feature Change BoardConfig border style to allow adding a Help menu that links to the .chm. [RH]
25-JUL-2016 Bug fix Constant current steps with no limits and only voltage ends would not run in hyperswitch mode.
26-JUL-2016 Bug fix Minimum value of AUXPingDelay is now 100 ms. Extended CAN frames did not work. Fixed.
28-JUL-2016 New feature Full admin rights no longer required.
29-JUL-2016 Bug fix Restart into a function step allowed except for EXTERNAL functions (DLLs).
01-AUG-2016 New feature Sub Routines added to hyper switch mode.
02-AUG-2016 New feature Handling of end conditions with +d1, -d1, +d2, -d2, >d1, <d1 operators migrated to software.
08-AUG-2016 Bug fix For 8500 (type 6) Voltage Simple Cal, check Vsense high-voltage upon starting cal. Save same
values to Chg and Dischg.[RH]
09-AUG-2016 Bug fix MacNet (7, 11) Fixed and specification slightly changed.
17-AUG-2016 Bug fix Init screens popping up during remote start fixed.
18-AUG-2016 New feature Minor version number incremented. Version is now: 3.2."build number"
More decimals added to the C-rate entry in initialization screen 1
Support up to 250 charaters in data file name, test comment, proc. name & proc. comment.
MacNet (6,2) Start test: add Start Data Type 2: Additional start parameters, Atomic start
23-AUG-2016 New feature Normal operation events can optionally be logged.
25-AUG-2016 New feature Optional text version of SysEvent.log added. WinEvent.txt size automatically limited.
14-SEP-2016 New feature Support for RabbitMQ message system added to MacNotify.
26-SEP-2016 Bug fix MacNet problem with ranging by power module fixed.
28-SEP-2016 Bug fix Fix uncombining channels not releasing slave channels. Allow Save Board on Ctrl Board Config
with suspended or completed channels. Add message to indicate that board was saved.[RH]
07-OCT-2016 Bug fix Rearrange layout of Check Channel form.[RH]
12-OCT-2016 Bug fix Hide ranges (on Check Channel form) not relevant to the board type or Ranging by Power
Module. Change combined channel status so that it updates without password entry.[RH]
13-OCT-2016 Bug fix Current ranges were not checked correctly in waveforms.
18-OCT-2016 New feature Preliminary support for v5_cntl1.bin added.
20-OCT-2016 New feature H2 and H3 bricks can now be combined.
24-OCT-2016 Bug fix Correction for relay switch timing error
27-OCT-2016 New feature Update lost heartbeat conversion from tics to seconds.
01-NOV-2016 Bug fix MacNet (6, 8) set direct mode output delay reduce to 10 ms for lab systems.
03-NOV-2016 New feature EXTERNAL, i.e. DLL call added to rest steps.
08-NOV-2016 Bug fix EXTERNAL, i.e. DLL call revision A added.
12-NOV-2016 New feature JSON protocol added to MacNet.
16-NOV-2016 Bug fix MacNet (7, 2) SMB suspend fixed.
29-NOV-2016 New feature Backup copies of SysEvents.log added.
14-DEC-2016 Bug fix Verification of board timing data updated.
22-DEC-2016 New feature AC impedance on large channels optionally allowed.
26-DEC-2016 Bug fix Revise Combined Channels. Changes correspond to Calibration changes.[RH]
03-JAN-2017 New feature EMLP added.
05-JAN-2017 Bug fix Update EV Type 31. Show Board Timing panel only when set in Debug.ini.[RH]
27-JAN-2017 New feature Cell Balancing option added.
08-FEB-2017 New feature More decimals added to capacity and energy in advanced start.
20-FEB-2017 New feature Additional timing logging added to MacNet
21-FEB-2017 New feature H bricks can now run on 1 channel controller boards.
03-MAR-2017 Bug fix Update Brd Type to match minorTic change in Cal program. Update Brd Config to match
settings changes in Cal: fix Board Type combo settings, remove obsolete Quad Range.[RH]
03-APR-2017 New feature Support for SystemType 7 added.
04-APR-2017 Bug fix Enable Board timing panel only when editing is allowed.
10-APR-2017 Bug fix Subroutine step following WF step couldn't start. Fixed.
18-APR-2017 New feature Call EVMaster.exe from EVChamberU when EV Chamber is type 1000 (representing multiple
types) in System.ini, and call MacEVcontroller.exe otherwise.[RH]
21-APR-2017 Bug fix MacNet: Reported current forced to 0 in non-active steps.
28-APR-2017 New feature Resistance pulsing permitted in "10 ms mode"
01-MAY-2017 Bug fix Add default relay times in Board config.[RH]
04-MAY-2017 Bug fix Specific values and rates are extracted and used from existing data file when restarting a test.
17-MAY-2017 New feature Setting RANDOM_As_Default_Name to 3 causes old random file name to be used.
18-MAY-2017 New feature Add EV Type 42 Binder RD3. [RH]
In SimpleCal, add ChMaint_GO ChMode 8 for Type 7 voltage cal.
29-MAY-2017 New feature Check and re-transmission of step in MacNet direct mode added.
31-MAY-2017 New feature Add EV Type 41 Oyo Electronics CLCD-9700-GF1.[RH]
01-JUN-2017 Bug fix Early H-Bricks in wrong range after rest steps plus current spikes fixed. (Firmware)
11-JUN-2017 Bug fix MacNet (1, 534), (1, 535) and (1, 536) are now only fired if the MacNet communication is set
to be logged.
22-JUN-2017 Bug fix Restructuring of program flow and communication.
28-JUN-2017 New feature Regimes are now supported with barcodes.
29-JUN-2017 Bug fix Update MaxChnsThisBrd array from size 12 to size 30 to accommodate comb. chns. on systems
with > 12 ctrl. boards.[RH]
02-JUL-2017 Bug fix Fix channels starting after test selected but before modifications are made on Inititialization
Screen 1.[RH]
24-JUL-2017 New feature More decimals added to C-rate in detailed view.
25-JUL-2017 Bug fix Init screen rearranged to be useful in 1024x768 resolution
26-JUL-2017 Bug fix "Constant current steps with no limits but with voltage ends, get this end set as a limit.
SystemType 0 and 3 only" is now optional and is by default OFF.
03-AUG-2017 New feature SMBF0X, SMB_READ_AGE, SMB_CLIENT_STAT and SMB_OK_0X added to Function data.
04-AUG-2017 Bug fix Pulse function values fixed
07-AUG-2017 Bug fix Check for too large end and report step times added.
10-AUG-2017 New feature Add system type 3, up to 8 combined channels.[RH]
11-AUG-2017 Bug fix Internal defaults for "long step times" changed. [JM]
False positive (1, 538) system events eliminated. [JM]
Fix uncombine of types 1,3.[RH]
14-AUG-2017 Bug fix v5_cntl1.bin is no longer auto-created as a copy of v5_cntl8.bin if absent.
15-AUG-2017 Bug fix v5_cntl8.bin is used for 1 channel controllers in hyper switch mode.
18-AUG-2017 Bug fix Deleting archived files stopped communication. Fixed. More regular controller board
communication during start of multiple channels to address long delays in MacNet.
22-AUG-2017 Bug fix Check added to ensure MacNet channel output (6, 8) is not set too soon after start (6, 7).
Could cause a controller board reset.
25-AUG-2017 Bug fix Minor tweaks to further improve MacNet throughput and stability.
29-AUG-2017 New feature Modulab interface updated to support Modulab version 3.3
31-AUG-2017 New feature VersaStudio interface updated to support version 2.52.3
11-SEP-2017 Bug fix Memory leak in cell balancing fixed.
03-OCT-2017 Bug fix No time reports and setting VARs could cause a very high rate of data points. Fixed.
Special_GUI=Stacklight-option-3 would revert to Maccor. Fixed.
06-OCT-2017 Bug fix Creation order of test procedure unit changed. Incomplete balance step (Occassionally, cell-
balance step gets cancelled prematurely just after start). May leave pack un-balanced.
20-OCT-2017 Bug fix Cosmetic update of function readings entry in BuildTest.
CellBalancer: Ensure all balancers in group get the balancing test command. Added a few
retries before indicating error.
27-OCT-2017 New feature Add EV type 43 "Binder Addressable".[RH]
31-OCT-2017 Bug fix M419: Capacity and energy entered in adv. start were ignored when restarting into function
chg and dchg steps. Fixed. [JM]
02-NOV-2017 Bug fix Add error handling for TCP EV controllers. Fix memory leaks in EVMaster.[RH]
03-NOV-2017 Bug fix M416: No CAN data handling after restarting SW, suspend-modify-resume, restarting a stopped
test and adv. start. Fixed.[JM]
06-NOV-2017 Bug fix M417: CAN: First signal not working in functions fixed [JM]
10-NOV-2017 New feature Support for FRA (currently Modulab) added for SystemType 3
Support for 32 channel FRA MUX added.
32 Channel Mux support added to ManualMux, MacIP, SetFRAMuxU, IPU [JO]
13-NOV-2017 Bug fix Fix Maint. Mode start channel not stopping when changing channels in Check Chan form.[RH]
21-NOV-2017 New feature Support for FRA (Solartron 1260/1287 and VersaSTAT (P4K)) added for SystemType 3 [JO]
23-NOV-2017 New feature DC bias option added for VersaSTAT / P4K FRA's. [JO]
27-NOV-2017 New feature DC bias option added for Modulab / Solartron 1287 [JO]
07-DEC-2017 Bug fix Restore sharing of Serial component using Environmental_Chamber_Use_COM_Port=1 (COM as
Address) for EV types 2,10-13,15,17. Add types 27,29,30,34,36,39. [RH]
15-DEC-2017 New feature Active channels are no longer automatically resumed after controller board reset. Can optionally
be turned on with the "AutoRestartAfterBrdReset" in the System.ini. [JM]
15-DEC-2017 Bug fix Fix I/O overlap issue on COM as Address type EV chambers.[RH]
18-DEC-2017 Bug fix Problems during waveform exit trapped and handled. [JM]
20-DEC-2017 Bug fix Report conditions handled incorrectly when SW reports are present. [JM]
20-DEC-2017 Bug fix M390: Fix Volt, Amp, Watt units not being retained after closing MacTest32 (all VB).[RH]
21-DEC-2017 Bug fix Erroneous EMLP checks on aux and SMB boards fixed. [JM]
07-JAN-2018 Bug fix Add limitation to Type 7 (VB) that channel cannot start if its voltage is already less than its
min. voltage spec.[RH]
08-JAN-2018 Bug fix Added MuxType to FRAForm {MuxType}. Add type TMuxes(CH08MUX, CH32MUX). Fixes
accessing > 1 mux of either type. [JO]
09-JAN-2018 Bug fix On Check Channel form, restrict hyperswitch to Range 4 chg/dis. Prevent Const. Voltage in
dchg. Hide range panels for ranges not used in single range and Ranging by Modules
(dual-range/triple-range/hyperswitch).
Close EVController if type31 stops communicating. [RH]
09-JAN-2018 Bug fix Problem with 8 reports in a step fixed. [JM]
12-JAN-2018 Bug fix Stop setting all boards' SWFlags to 2 when System.ini has Use Board Timing (GetBrdConfig VB).
Relay timing should be controlled only by Use Board Timing checkbox in Calibration.[RH]
17-JAN-2018 New feature MacNotify now tolerates trailing spaces in SystemID. [JM]
30-JAN-2018 New feature Mux 4600M now compatible with MacTest32 in fixed IP board config. Use_First_Address as last
byte of Address and MAC. [JO]
30-JAN-2018 Bug fix Prolonged Start-Screen at 22%. Issue in MacIP. (Fixed.) [JO]
31-JAN-2018 New feature Add alternate binaries for other 8500 (type 7) configurations.[RH]
01-FEB-2018 Bug fix Starting cycle and number of cycles fixed. [JM]
08-FEB-2018 Bug fix Reduce lag time communicating with 8500's (type 7) by setting PollMax in MainPollTester to 1
for 8500's.[JM/RH]
19-FEB-2018 Bug fix "Step2X" problem fixed in firmware. (KR)
03-MAR-2018 New feature VersaSTATwrapper.dll: Enable current limits on ParSTAT 4K.
ParSTAT4K: Enabled choice between OpenCircuit and Reference [JO]
05-MAR-2018 New feature SMBF0X in functions can now read up to 8 bytes of string data. [JM]
06-MAR-2018 New feature Support for hexadecimal numbers added to functions. [JM]
08-MAR-2018 Bug fix 500 ms only: Prevent premature current ends in voltage limited steps. [JM]
22-MAR-2018 Bug fix Check for sub routines with infinite loops added. [JM]
28-MAR-2018 Bug fix Failing aux and smb cause assigned channels to enter a problem state.
Crash of Susp-Mod-Resume corrected. [JM]
16-APR-2018 Bug fix Failing CAN cause assigned channel to enter a problem state. [JM]
01-MAY-2018 Bug fix Quick wiev charts fixed. [JM]
08-MAY-2018 Bug fix MacQCShell.exe could not start tests. Fixed. [JM]
10-MAY-2018 New feature Add AES XChange Board EVType 44. [RH]
22-MAY-2018 New feature FileRemCtrl is now an option that is off by default. [JM]
30-MAY-2018 Bug fix Data files are kept open to avoid lock ups. [JM]
05-JUN-2018 New feature Updated ManualMux to control all muxes on the tester (ManualMux, SetFRAMuxU) [JO]
11-JUN-2018 New feature Allow Modulab tests written inside XM studio can run through MacTest32. Data is written back
to test path. [JO]
19-JUN-2018 Bug fix Modulab machine prefers shorter file names with no spaces. Also has trouble running multiple
chns at a time. Changed from using FRAStatus to FRAQueue. (FRAU, ModuLabU) [JO]
19-JUN-2018 Bug fix Waveform file names of numeric characters only failed to run. Fixed. [JM]
21-JUN-2018 New feature Add maintenance mode for Digital output testing and development (positions A0-7 only). [RH]
02-JUL-2018 New feature Profiling added [JM]
06-JUL-2018 Bug fix Intermediate and normal screen updates improved [JM]
10-JUL-2018 Bug fix Fix MacIP struggling to manage concurrent multiple DHCP address requests by adding queue to
manage requests.(IPMainU) [JO]
19-JUL-2018 New feature Updated VersaSTATwrapper.dll to expose more functions
20-JUL-2018 Bug fix Applied fix for ParSTAT/VersaSTAT selecting wrong current range when set to AUTO.
Updated VersaSTATwrapper.dll file. This DLL must be packaged with any SW build for proper
function of FRA in MacTest32. (FRAU, ParSTAT4000U, VersaSTATwrapper.dll)
01-AUG-2018 Bug fix Tests entering a sub routine caused erroneous start of other chns shown in Init.screen 1. [JM].
08-AUG-2018 Bug fix ModuLab showing Potentiostatic instead of Galvanostatic during FRA test. Fixed [JO]
09-AUG-2018 Bug fix ChannelNumberOffset was not applied in EV details.
16-AUG-2018 Bug fix Manual-Mux testing won't go past the 2nd board. Fixed [JO]
07-SEP-2018 New feature Allow SimpleCal for ECC/ELC. [RH]
18-SEP-2018 Bug fix FRA compatibility with 50ms system (4600M). Mostly cleanups.
Remove support for obsolete Solartron 1470.
28-SEP-2018 Bug fix Multiple fixes of password controlled access [JM]
28-SEP-2018 Bug fix Cellbalancer occassionally misses data points. Approach to collecting and logging data
re-visited. In case of a missed data point, use last data point [JO]
02-OCT-2018 New feature CAN default addressing. Assigns address up to 192.168.127.244
03-OCT-2018 Bug fix Improved set up of the first user [JM]
SW hangs/crashes when CAN is enabled but device(s) unresponsive. Fixed by ignoring
StartCAN if bad comm is found on any CAN device. Improve MacIP servicing of addresses.[JO]
04-OCT-2018 New feature Added option to terminate MacIP when MacTest32 closes (Settings.ini).[JO]
05-OCT-2018 Bug fix More refinements of password controlled access [JM]
05-OCT-2018 Release Version 3.3.0.92
12-OCT-2018 New feature Option for ignoring failing SMB channels added.
23-OCT-2018 Bug fix Option for ignoring failing SMB channels extended.
30-OCT-2018 New feature Add Report Problems button under Maintenance. [JO]
08-NOV-2018 New feature Add Espec P310 EVType45. [RH]
09-NOV-2018 Bug fix Avoid port exhaustion on EVType31 by increasing query interval to reduce Ports opened.[RH]
15-NOV-2018 Bug fix Function steps with constant values, did not always set the right value after resume. Fixed[JM]
16-NOV-2018 New feature Add EVType44 power fail feature. [RH]
17-DEC-2018 New feature Password-protect Calibrate Aux button on Assign Aux. Inputs form (under Maintenance). [RH]
19-DEC-2018 Bug fix Potential time cogging with WF fixed. [JM]
20-DEC-2018 New feature MacNet (1, 2) extended with Channel Number Offset.
MacNet (4, 11) File name, test proc. name and comments extended.[JM]
08-JAN-2019 New feature Function Detail form size is now variable.
10-JAN-2019 Bug fix FileRemCtrl is now always active for SystemType 1. [JM]
11-JAN-2019 New feature Set EV Temp report added [JM]
14-JAN-2019 Bug fix Fix negative temperature handling on EV Type 41. [RH]
23-JAN-2019 New feature <pk operator execution moved to SW. [JM]
24-JAN-2019 Bug fix Occasional "Error 9" when starting systems with channel offsets resolved. [JM]
30-JAN-2019 New feature Preliminary SMB Charge step added [JM]
31-JAN-2019 Bug fix Crashing during shut down from Formation Shell corrected. SMB details dialog appearance
adjusted. Issue detecting failing SMB channels addressed. [JM]
07-FEB-2019 Bug fix EV steps immediately ending steps following steps on busy systems addressed [JM]
14-FEB-2019 Bug fix False reports during set EV Temp report removed [JM]
15-FEB-2019 New feature Add Votsch ASCII-2 J-Bus as EV Type 46 [RH].
21-FEB-2019 New feature SMB Charge step added limits [JM]
Type 7: enable Reset Bd, revise Simple Cal display, utilize DAC cal for P & I scale values.[RH]
22-FEB-2019 Bug fix False error in Set EV Temp report addressed [JM]
28-FEB-2019 Bug fix Threshold time increased to prevent false reporting of system event (0, 206) [JM]
28-FEB-2019 New feature Add Lauda Proline EV Type 47 [RH].
13-MAR-2019 Bug fix AC-impedance on combined channels prevented. Dump folder obsoleted. [JM]
14-MAR-2019 Bug fix AutoRestartAfterBrdReset hidden. [JM]
14-MAR-2019 Release Version 3.3.114
18-MAR-2019 New feature Obsolete ServiceSettings removed [JM]
04-APR-2019 Bug fix For ParSTAT4000, undo Aug. 2018 change dividing FRA ampl./10 in potentiostatic mode. [RH]
05-APR-2019 Bug fix Update Power Fail for EVType44. [RH]
08-APR-2019 Bug fix Handling of locked WF files added [JM]
09-APR-2019 Bug fix Problem with Brick power safety addressed [JM]
11-APR-2019 Bug fix Displayed discharge brick power values in maintenance screen was wrong. Fixed. [JM]
12-APR-2019 Bug fix Assigned non-responding SMB would cause maintenance mode to stop. Fixed. [JM]
22-APR-2019 New feature Add 8500 Auxiliaries. Update SysEvents for 8500 Aux's also. [RH for JO]
24-APR-2019 New feature EMLP permitted on H-bricks. [JM]
26-APR-2019 Bug fix Barcode details addressed. (M38) [JM]
07-MAY-2019 New feature Add Tools>8500 Auxiliary Details in VB. Add Aux8500Pgm unit [RH for JO]
15-MAY-2019 New feature Regime support added to MacNet function (6, 2) [JM]
16-MAY-2019 Bug fix Legal length of LOT no in Init Screen 5 corrected.
Waveforms no longer generate false (0, 215) events. [JM]
24-MAY-2019 New feature Support for Npcap added [JM]
04-MAY-2019 Bug fix SMB event (1, 451) excluded from (0, 226) check.
30-JUN-2019 Bug fix Hole in regime start checks mended. [JM]
30-JUN-2019 Release Version 3.3.130
23-JUL-2019 Bug fix Fix EVType 47 communication [RH]
29-JUL-2019 Bug fix SMB Write no longer generate false (0, 215) events. [JM]
12-AUG-2019 Bug fix Digital output steps no longer generate false (0, 215) events. [JM]
13-AUG-2019 Bug fix Board reset when leaving multi level subroutines caused channels to go available (M56) [JM]
16-AUG-2019 New feature Update to BasicLAN for customer shell program. [RH]
21-AUG-2019 Bug fix Update to resolve "lost focus" issue when using barcode scanner (M38) [RH]
30-AUG-2019 Bug fix MacNet (6,2) Start Test: Report problems back instead of displaying on the screen. [JM]
31-AUG-2019 New feature Service MacNet msg type (4, 30) to allow passing auxiliary digital input for shell.[RH]
04-SEP-2019 New feature Service MacNet msg type (6, 2:10) to set all channel safeties for shell.[RH]
10-SEP-2019 Bug fix Resuming 96 channels in pause steps causes the SW to lock up addressed (M55) [JM]
23-SEP-2019 Bug fix H3 brks only: Convert Rest before sub routine to rng. 1 dchg current 0 to prevent lock ups.[JM]
24-SEP-2019 Bug fix JSON version of MacNet (3, 1) Channel specification caused a lock up. Addressed [JM]
03-OCT-2019 New feature Optional set function in different step added [JM]
07-OCT-2019 New feature Add Terchy S8521 EVType 48. [RH]
08-OCT-2019 Bug fix Starting a test with entry and exit of subroutine no longer generate false (0,220) events. [JM]
17-OCT-2019 Bug fix Fix number of chambers running as Type 45. [RH]
21-OCT-2019 New feature Adjustment of thermocouple and thermistor filtering added [JM]
21-OCT-2019 Bug fix Crashing during power fail exit rectified [JM]
28-OCT-2019 Bug fix H3 combined only: Convert Rest before Dchg to rng. 1 dchg current 0 to prevent lock ups [JM]
29-OCT-2019 Bug fix H3 combined only: Convert Rest before Chg to range 1 chg current 0 to prevent lock ups. [JM]
11-NOV-2019 Bug fix No Reports in waveforms optimized for lower data rate. [JM]
20-NOV-2019 New feature Add handling for CRate in CheckStartTest (for MacNet and 4000Shell). [RH]
09-DEC-2019 Bug fix Add missing support for 15k thermistor panel type.[RH]
10-DEC-2019 Bug fix Misc adjustments for more stable waveform [JM]
10-DEC-2019 New feature Max value of AiThermoFilter increased to 40 [JM]
16-DEC-2019 Bug fix MacNet (1, 5) file listing problem with empty lists fixed. [JM]
12-DEC-2019 Bug fix Waveform overrun (congestion) protection [JM]
17-DEC-2019 Bug fix SW generated ends are now delayed 2 ticks into steps to address
problems with delta operators working incorrectly [JM]
20-DEC-2019 New feature Check of waveform block length added. [JM]
02-JAN-2020 Bug fix Error in +d2 and -d2 end operators fixed. [JM]
07-JAN-2020 New feature Negative delta X permitted for differential operators [JM]
13-JAN-2020 Bug fix Start test from file didn't work with ChannelNumberOffset enabled [JM]
21-JAN-2020 Bug fix Incorrect trip of system event (1, 171) [JM]
Start test with file did not work if channel 1 was in a subroutine [JM]
22-JAN-2020 Release Version 3.3.143
29-JAN-2020 Bug fix Redundant Voltage Safety check in SW suspended [JM]
Incorrect status messages from relay cycle ignored [JM]
Occasional false Safety Shut Down fixed [JM]
02-FEB-2020 New feature Add EVType 49 KSON THS-600. [RH]
11-FEB-2020 New feature Waveform version 1.1. Header added for faster access [JM]
16-FEB-2020 Bug fix Only actually used ends and reports sent in waveforms [JM]
21-FEB-2020 Bug fix Clear SMB write buffer if channel becomes not active [JM]
25-FEB-2020 Bug fix Further optimization of waveform download and optimization [JM]
New feature "X" report type added to waveforms [JM]
28-FEB-2020 New feature Extend combined channels management from 30 controllers to 64.[RH]
10-MAR-2020 New feature Update BasicLAN for type 7 8500. [RH]
23-MAR-2020 Bug fix Multiple improvements of waveform throughput [JM]
26-MAR-2020 New feature Calls MIMS Client 2 and View Data 2
26-MAR-2020 Release Version 3.3.148
30-MAR-2020 New feature Allow resume Digital Outputs option. [RH] (Mantis 48)
07-APR-2020 Bug fix % of Max Data corrected in SMB setup
08-APR-2020 New feature SMB 3.1
15-APR-2020 New feature MacNet (7, 24) added. Refresh SMB scan list [JM]
13-MAY-2020 New feature Device addressed added to "SMB Val" end condition and "Set SMB" report
02-JUN-2020 New feature Support for 1-byte SMB registers added
16-JUN-2020 New feature Support for 20A 12V Solartron EnergyLab XM added.[RH]
17-JUN-2020 Bug fix Multiple improvements and fixes addressing high rate waveforms on many channels
At present just for 2-8 channel per board lab mode [JM]
18-JUN-2020 Bug fix 20A 12V booster was not recognized [RH]
12-AUG-2020 Bug fix Recovery from power would occasionally cause a loaded DLL to disconnect [JM]
14-AUG-2020 Bug fix Multiple improvements and fixes addressing high rate waveforms on many channels
At present just for 2-8 channel per board lab mode [JM]
27-AUG-2020 Bug fix Further clean up of waveform fixes. [JM]
09-SEP-2020 New Feature Company name added [JM]
21-SEP-2020 Bug fix Regime start congestion fixed [JM]
29-SEP-2020 Release Version 3.3.167
30-OCT-2020 New Feature Add Watlow F4T with profile engine-based start/stop (EV51). [RH]
02-NOV-2020 Bug fix Faster release of socket in Windows after closure.
11-NOV-2020 Bug fix Fix Multi-chamber bug introduced in March.[RH]
04-DEC-2020 Update Security update
21-JAN-2021 New feature Add Huber PilotOne (EV Type 53). [SL]
21-JAN-2021 Bug fix Power_Fail_Restart_Time of 0 failing when power fail < 1 sec.(Mantis 86)[RH]
25-JAN-2021 New feature MacNet(4, 12) added for active step info.[RH]
16-MAR-2021 Bug fix Waveform problems for single channel lab and 8500 addressed [JM]
17-MAR-2021 Bug Fix Incorrect reporting of (0, 73) calibration error for 8500 corrected [JM]
17-MAR-2021 New feature Add support for Maccor multi-chan Ethernet-CAN board.[RH for JO]
18-MAR-2021 Release Version 3.3.168
14-JUN-2021 Bug Fix Correct swapped JSON pair labels SMBRegAddr/SMBDevAddr for SMB (7,4) query.[M109]
10-AUG-2021 Bug fix Fix crash when using Adv. Start with large system.(Mantis 126)[RH]
31-AUG-2021 Bug fix Fix inappropriate limit enforcement on MacNet Direct Mode.(Mantis 153)[RH]
15-SEP-2021 Bug fix Channels leaving EV steps causing timeouts fixed [JM]
18-NOV-2021 New feature Add 8500 Verify. [RH]
20-DEC-2021 New Feature SMB4 support added [JM]
04-JAN-2022 Bug Fix VARs and Global flags were not updated in data files (M244) [JM]
15-FEB-2022 Bug Fix Incorrect double-quotes in MacNet JSON error messages fixed. [JM]
16-FEB-2022 New Feature External DLL RevB added including SMB4 support. [JM]
18-FEB-2022 New Feature SMB4 support added to functions. [JM]
07-MAR-2022 New Feature Add EVType 61.[RH]
30-MAR-2022 New Feature MacNet (6, 14) Set channel start data added [JM]
03-MAY-2022 Bug Fix Fix Big-Endian handling for CAN.[RH]
04-MAY-2022 New Feature Add EVType 62.[RH]
17-MAY-2022 Bug Fix Correct handling of extended CAN IDs above 65535.[RH]
15-JUN-2022 New Feature Increased update rate of external DLL's when reading SMB4 [JM]
27-JUL-2022 Bug Fix Handle occasional problem with EventsRateThreshold reporting 0.[JM]
28-JUL-2022 Bug Fix Erroneous call of event (0, 226) addressed [JM]
16-SEP-2022 New Feature MacNet (7, 12) optionally extended with timing info [JM]
07-NOV-2022 Bug Fix MacNet (7, 14) returned obsolete board device address. Removed [JM]
07-NOV-2022 Bug Fix Client status under SMB4 details GUI is not updated correctly after clicking refresh button. Fixed (M333) [JM]
08-NOV-2022 New Feature Support I2C option in Set SMB and Set SMB Ext from test procedure added (M351) [JM]
10-NOV-2022 Bug Fix Sporadic bad SMB4 reads addressed (M338) [JM]
10-NOV-2022 Bug Fix Return data for Macnet SMB (7,2) - Suspend and (7,3) - Resume added (M344) [JM]
12-DEC-2022 Bug Fix EV-CAN fixes: missing EV-CAN data from SubR, datafile too large, chamber to match chn [RH]
15-DEC-2022 Bug Fix Misc SMB4 updates including client FW download.
02-JAN-2023 Bug Fix SMB4 bootloader events added.
04-JAN-2023 Bug Fix Special step type for Cell Balance on 8500.[RH]
06-JAN-2023 Bug Fix Enable "Force Step" to shut down Cell Balance step more promptly.[MG]
09-JAN-2023 Bug Fix Support for Smart Battery Readings (SMB4) Boards tab.(M371)[RH]
19-JAN-2023 New feature Add Suspend, External Control, Adv. Start for EV Type 62. [RH]
07-FEB-2023 New Feature SMB4 v2 FW added [JM]
10-FEB-2023 Bug Fix SMB4 GUI is not updated correctly after disconnect the pack (M340) [JM]
13-FEB-2023 Bug Fix SMB4 Get Header does not get header data (M370) [JM]
Sometimes scanlist get crashed (M380) [JM]
SMB4 may lost scanlist for few positions under reading (M331) [JM]
Lost of write commands (M376) [JM]
13-FEB-2023 Bug Fix 29-JAN-2020 Bug fix Incorrect status messages from relay cycle ignored. Caused missing end checks for formation. [JM]
Ping rate too low in 50 and 500 ms systems addressed.
14-FEB-2023 New Feature String-I2C option added to SMB4 (M348) [JM]
20-FEB-2023 Bug Fix (M213) addressed by adding a "No report EXT" (Y), which works as "Only 1 report" but without addding data to the file [JM]
23-FEB-2023 SW Fix Common device address added for MacNet (7, 14) in SMB4 (M374)
MacNet and JSON (7,1) Get SMB info will lose comm with SMB3 on legacy fixed (M350)
Boards tab under SMB4 detail screen completed (M371)
23-FEB-2023 SW Fix Device address added for MacNet (7, 14) in SMB4 as the address in the first scan list entry (M374)
07-MAR-2023 SW Fix Suspend/Resume of non-dll step triggers OnSuspend/OnResum fixed (M403) [JM]
08-MAR-2023 SW Fix Fixed CellBalancing for groups of balancers divided amongst channels [MG]
13-MAR-2023 SW Fix Convert Rest before sub routine and WF to same type and range as the sub and WF. (M230) [JM]
28-MAR-2023 New Feature Support for SMB4 added to user DLL RevA [JM]
11-Apr-2023 SW Fix Undesired StepEnd and StepStart events on Suspend Resume fixed (M403) [JM]
14-APR-2023 SW Fix Adjustment SMB write with SMB4 HW from user DLL RevA (M418) [JM]
21-APR-2023 SW Fix Multiple fixes of External DLL events [JM]
Prevent accidental refresh of all SMB4 scanlists. [RH]
01-MAY-2023 SW Fix Range change in 1 tick steps prevented (M330) [JM]
02-MAY-2023 SWFix V limit from external DLL gets ignored fixed (M286) [JM]
Dlls don't work after Maccor was forced closed fixed (M214) [JM]
03-MAY-2023 SW Fix LostSMBOption activated for SMB4 [JM]
04-MAY-2023 SW Fix Brick Power Safety eliminated (M191) [JM]
14-MAY-2023 SW Fix Make I2COption default on in MacNet (7,10) - SMB write (High bit: No Option = 0 = I2CWrite, 1 = SMBWrite) [MG]
15-MAY-2023 SW Fix Changed ExecuteChanLocal so Event(7,7) can only be reported once per second. This should also prevent force step from being executed twice[MG]
18-MAY-2023 SW Fix Reversed May 14th change. MacNet (7,10) reverted to original behavior for UDP/TCP. (High bit: No Option=0=SMBWrite, 1=I2CWrite)[MG]
24-MAY-2023 SW Fix SMB4 Val end type not working fixed (M442) [JM]
24-MAY-2023 SW Fix Changed SMB4->SMB3 Copy Memory to correct LegacyBlock data shift (M439). [MG]
24-MAY-2023 SW Fix Correct length of MacNet (4, 12) [RH]
25-MAY-2023 SW Fix Added SMB4 data to MacNet (1,2) for both UDP/TCP and JSON (M445) [MG]
31-MAY-2023 SW Fix OnStepStart fired before OnUnload when channel is reset fixed (M440) [JM]
01-JUN-2023 SW Fix Reduce command burden on controller during Jump To Step that caused No Controller. (M438) [RH]
03-JUL-2023 SW Fix Waveform and error checking optimization [JM]
05-JUL-2023 New Feature CAN code Modified for data priority and general MacTest32 responsiveness. MacTest32 can now support more CAN-devices (more short-interval frames) [JO]
with little to no drag/unresponsiveness. Update keeps CAN data close to real-time as possible (very important for high volume frames, especially in [JO]
multiple CAN-devices environment with high frames characteristics); important for customer diagnostics, especially in case of failure. [JO]
Added a window (CANMainMsgU) to notify of device's status and notify of communication activities and error messages. MacTest32 does not need a [JO]
re-start. if CAN-device(s) goes offline temporarily, or was offline at start/running; all CAN-devices will be found and polled. [JO]
Affected Files {CANMainU.pas/.dfm, AuditU.pas, CANHandlerU.pas} [JO]
06-JUL-2023 SW Fix SystemEvent 1, 152 made optional [JM]
31-JUL-2023 New Feature Configurable minimum time between reception of a network message to transmitting the following added [JM]
17-AUG-2023 New Feature Non-blocking download of V5_cntl8.BIN
18-AUG-2023 New Feature Handling of network messages received in the reverse order [JM]
New Feature Optional AutoResumeAfterBoardReset added [JM]
22-AUG-2023 New Feature Warning for both NPCap and WinPCap being present added [JM]
11-SEP-2023 New Feature Optional AutoResumeAfterBoardReset misc fixes [JM]
14-SEP-2023 SW Fix EventsRateThreshold defualt changed from 200 to 1000 (M473) [JM]
19-SEP-2023 SW Fix VersaStat VDK updated to version 2.66.1.0
Bad DAC calibration check
28-SEP-2023 SW Fix Avoid scientific notation numbers in PartSTAT parameters [JM]
Erroneous parameter in PartSTAT galvanostatic mode removed [JM]
Download Version: 12-Oct-2023
MIMS Information Management System Version 1.0
Download 22 Jan 2020
Content for tab 3.